PGEE11213 Chap.2 Error, Uncertainty and Measurement Decisions
Error, Uncertainty and Measurement Decisions
Keep error, correction and uncertainty separate
Error is the difference between an indication and a reference value, although the exact error is generally unknown. A correction compensates for a recognised systematic effect. Uncertainty describes the dispersion of values reasonably attributable to the measurand.
These ideas require different actions, so using one word for all three hides the measurement model.
Combine standard contributions through the model
Express each contribution as a standard uncertainty, apply its sensitivity coefficient and include covariance when inputs are dependent. Independent terms combine by root-sum-square. Expanded uncertainty multiplies the combined value by a stated coverage factor.
The budget should reveal which contribution dominates and where process improvement will matter.
Make the conformity rule explicit
Tolerance limits describe acceptable product geometry, while the uncertainty interval qualifies knowledge of the result. When that interval crosses a specification limit, the central estimate cannot settle conformity by itself.
Apply an agreed decision rule or guard band and state the resulting risk rather than silently ignoring uncertainty.
What this chapter covers
- 01
Error and correction
- 02
Type A evaluation
- 03
Type B evaluation
- 04
Sensitivity coefficients
- 05
Combined uncertainty
- 06
Conformity decisions
Worked example · free
Combine an uncertainty budget
- 2Square the three standard contributions and add them.
- 2Take the square root to obtain 0.0153 mm combined uncertainty.
- 2Multiply by 2 and report expanded uncertainty as 0.031 mm.
Key terms
- Correction
- A Correction is an estimated adjustment applied to compensate for a recognised systematic effect.
- Combined Uncertainty
- Combined Uncertainty propagates standard contributions through the measurement model into one standard uncertainty.
- Guard Band
- A Guard Band narrows an acceptance region to control decision risk when measurement uncertainty is material.
Error, Uncertainty and Measurement Decisions FAQ
Why are independent uncertainties combined by root-sum-square?
Independent standard contributions represent dispersion from separate mechanisms, so their variances add through the measurement model. Taking the square root returns the result to the unit of the measurand. Direct addition is a different conservative rule and should not be used without stating why it fits the decision.
What should happen when an uncertainty interval crosses a limit?
Report the estimate and expanded uncertainty, then apply the agreed conformity rule. The next action may be a guard band, improved measurement, rework or documented risk decision. Repeating the same setup only helps when short-term scatter is important; it does not remove calibration, alignment or model uncertainty.
How can an uncertainty budget improve a process?
Compare the squared contribution of each standard term to the total and target the mechanism with the largest leverage. Recompute the budget after a realistic improvement to estimate benefit. Reducing an already small term may be easy but will barely change the combined result or the conformity decision.
Why must the coverage factor be reported?
Expanded uncertainty cannot be interpreted from the numerical width alone because the coverage factor links the combined standard uncertainty to an intended level of coverage. A statement such as U = 0.12 mm is incomplete if it omits the factor, assumptions and units. Record the combined standard uncertainty, selected coverage factor, resulting expanded uncertainty and the confidence convention or distribution basis.
When effective degrees of freedom or a non-normal model changes the factor, preserve that calculation too. This lets a reviewer compare results consistently and prevents the expanded interval from being mistaken for a guaranteed error bound.
Assessment move
Rebuild one uncertainty budget without a spreadsheet. Write the measurement model, convert each input to a standard uncertainty, apply sensitivity coefficients and combine the terms. Then halve the dominant contribution and recompute the result. Finish with a conformity case whose uncertainty interval crosses a limit, and write the exact decision rule needed rather than declaring a pass or fail from the central estimate.
Make a source column for every contribution and state why its distribution and divisor apply. Change one rectangular contribution to a normal contribution with the same quoted half-width and observe how the budget changes; this exposes why a memorised divisor is unsafe. Work another example with a sensitivity coefficient that carries units, then check dimensional consistency before calculating.
Create three conformity diagrams: an interval fully inside limits, one crossing a limit and one fully outside. For each, write the action under a conservative rule and explain the producer or consumer risk. Finally, choose the largest squared contribution and propose a physical experiment that separates its likely causes. Update the budget only after the experiment produces evidence.
Keep the original and revised versions together so the claimed improvement can be reconstructed rather than inferred from a smaller final number. Extend the model with two correlated inputs and explain why adding their variances as independent terms can understate or overstate the result. You do not need a complex covariance example to learn the principle: identify the shared physical cause and predict the sign of association.
Next, audit rounding by reporting the uncertainty with sensible significant figures and matching the estimate decimal place. Compare that result with a version rounded at every intermediate step. Finish by explaining the budget aloud without reading labels. If you cannot identify the physical mechanism behind a row, the row is not yet understood well enough to defend in coursework or process review.