Chapter 5 of 8 · CHEM6118

Mass Analysers and Resolution

Different observables encode the same coordinate

Mass analysers separate ions by converting m/z into arrival time, trajectory stability or oscillation frequency. Time-of-flight instruments accelerate a packet and let lighter m/z values arrive sooner. Reflectrons and delayed extraction correct energy or position spread.

Quadrupoles transmit only trajectories stable under combined direct-current and radiofrequency fields, while ion traps confine and eject selected populations over time.

Compare specifications as independent axes

Resolving power measures separation, mass accuracy measures closeness to the true value, and precision measures repeatability.

Range, scan speed, transmission, isolation and source compatibility add further trade-offs. An analyser choice should name the analytical objective and the non-ideal spread that a design feature controls.

Match each analyser to its measured motion

In time of flight, equal acceleration gives ions comparable kinetic energy, so lower m/z produces higher velocity and earlier arrival.

A reflectron compensates partly for initial kinetic-energy spread by making faster ions travel deeper into an electrostatic field before turning. Delayed extraction addresses differences in starting position and energy before the flight begins.

Neither feature changes the identity of the coordinate; each reduces a source of arrival-time broadening.

A quadrupole instead filters trajectories in combined direct-current and radiofrequency fields. Only a selected stability region passes to the detector at a given voltage ratio. A trap applies related field principles over time, isolating, storing, activating or ejecting populations sequentially.

Ion cyclotron resonance measures a frequency in a magnetic field and can provide very high resolving power, but its performance comes with demanding fields, vacuum and acquisition requirements.

State the physical observable before comparing headline specifications.

Keep three quality claims separate

Resolving power asks whether neighbouring peaks can be distinguished and is commonly written as mass divided by a defined peak-width or peak-separation quantity. Accuracy asks how close a measured value lies to the accepted value, often in parts per million.

Precision asks how repeatably the measurement can be reproduced. A narrow but systematically shifted peak can be precise and well resolved yet inaccurate. An instrument decision should also consider mass range, scan speed, ion transmission, tandem capability and compatibility with the chosen source. Name which compromise matters for the analytical question.

In this chapter

What this chapter covers

  • 01

    Time-of-flight energy and arrival time

  • 02

    Reflectron energy focusing

  • 03

    Delayed extraction and orthogonal acceleration

  • 04

    Quadrupole trajectory stability

  • 05

    Three-dimensional and linear ion traps

  • 06

    Ion cyclotron resonance frequency

  • 07

    Resolution, accuracy and precision

Worked example · free

Calculate a minimum resolving power

Q [8 marks]. This AskSia-authored problem uses marks to guide study time only; it is not a University marking scheme. Two ions near m/z 750 differ by 0.050 u. Calculate the minimum resolving power under R = m divided by delta m and explain what it does not establish.
Time of flight converts equal acceleration into arrival-time separation.Time of flight converts equal acceleration into arrival-time separation. The blue lighter ion travels faster after receiving the same charge-dependent kinetic energy, so it reaches the detector before the heavier orange ion.Extractionsame zeVlighter m/zheavier m/zDetectorequal kinetic energy → lighter ions arrive first
Time of flight converts equal acceleration into arrival-time separation. The blue lighter ion travels faster after receiving the same charge-dependent kinetic energy, so it reaches the detector before the heavier orange ion.
  • 4Substitute mass and peak separation into the resolving-power relationship.
  • 4Distinguish separation from mass accuracy and practical baseline resolution.
R = 750 divided by 0.050 = 15,000. This is the minimum separation requirement under the stated convention. It does not establish that measured centroids are accurate, and an instrument operating exactly at that specification may provide little practical margin for peak shape or noise.
Sia tip — After calculating R, name the peak-separation convention and keep mass accuracy as a separate claim.
Glossary

Key terms

Time of Flight
A mass-analysis principle that separates accelerated ions by their travel time through a field-free region.
Reflectron
An electrostatic mirror that improves time focusing by compensating for part of the initial kinetic-energy spread.
Quadrupole
A four-rod analyser that transmits ions with stable trajectories under combined direct-current and radiofrequency fields.
Ion Trap
An analyser that confines ions and performs isolation, activation or mass-selective ejection over time.
Resolving Power
A measure of the ability to distinguish nearby peaks, commonly written as m divided by delta m.
FAQ

Mass Analysers and Resolution FAQ

Why do lighter ions arrive first in TOF?

Ions with the same charge accelerated through the same potential receive the same kinetic energy. Since velocity increases as mass decreases under that condition, lower m/z ions cross the flight tube sooner than higher m/z ions.

What problem does a reflectron solve?

Ions with the same m/z can begin with different kinetic energies and would otherwise arrive at different times. More energetic ions penetrate farther into the electrostatic mirror, increasing their path so they refocus near the detector.

Is higher resolving power always the best choice?

No. A method also needs suitable mass range, speed, transmission, source compatibility, isolation and cost. The best analyser provides enough separation for the actual chemical question while preserving the signal and time scale required by the experiment.

Study strategy

Exam move

For each analyser, write the measured difference and the equation linking it to m/z. Then add one source compatibility, one strength and one cost. Work resolving-power calculations separately from accuracy questions. For TOF, redraw how reflectron and delayed extraction correct different initial spreads.

Working through Mass Analysers and Resolution in CHEM6118? Sia is AskSia’s AI Chemistry tutor — ask any CHEM6118 Mass Analysers and Resolution question and get a clear, step-by-step explanation grounded in how CHEM6118 is taught and assessed. Read this chapter free, then take your hardest questions to Sia.

A+Everything unlocked
Unlocks this Bible + all 31 of your subjects at City University of Hong Kong, plus 1,000+ more Bibles across the AskSia library.
Sia - your CHEM6118 tutor, unlimited, worked the way the exam marks it
The full 4-page Bible + practice bank with worked solutions
Chrome extension - sync your LMS so Sia knows your deadlines
Bilingual EN / Chinese on every Bible and every Sia answer
$0.99 Trial
30-day money-back · cancel in one tap · how it works