CHEM6118 Chap.4 Molecular Ions, Isotopes and Ionisation
Molecular Ions, Isotopes and Ionisation
The ion source defines the population you observe
Mass spectrometry creates ions, separates them by mass-to-charge ratio and detects their abundance. Ion notation and charge state must therefore be established before converting m/z into neutral mass.
Electron ionisation often provides extensive fragmentation from volatile analytes, while electrospray preserves polar solution-phase molecules and commonly creates multiple charge states.
MALDI uses a matrix and pulsed laser desorption, often producing large singly charged ions.
Two independent routes to composition
Exact mass distinguishes formulas that share nominal mass when resolving power and accuracy are sufficient. Isotope envelopes test elemental composition through predictable M plus one and M plus two patterns.
Charge can be inferred from isotope spacing, but adduct masses must be removed explicitly. Agreement between ion-source chemistry, exact mass and isotope pattern is stronger than repeating one calculation.
Write the ion before solving for the molecule
An analyser reports mass-to-charge ratio, not neutral molecular mass.
Begin with explicit notation such as a radical cation, a protonated molecule or a multiply protonated ion. Multiply the observed coordinate by charge magnitude, then subtract or add the masses of the stated adducts to recover a neutral mass. Isotope spacing offers an independent charge check: neighbouring isotopologues are separated by approximately one divided by the charge magnitude in m/z.
A half-unit spacing therefore points to charge two rather than a half-unit isotope mass.
Source choice determines which ions are plausible. Electron ionisation suits volatile molecules and often creates reproducible radical-cation fragmentation. Electrospray transfers solution-phase ions and commonly produces protonated or deprotonated, multiply charged species, so solvent composition and salts can alter adducts.
MALDI couples pulsed laser energy to a matrix and often yields large singly charged ions. A missing molecular ion under a hard source does not establish that the neutral compound lacks the corresponding mass.
Cross-check formula evidence
Nominal mass uses integer mass numbers, exact mass uses the masses of specified isotopes, and average mass reflects natural-abundance weighting.
High resolving power can separate close formulas, but mass accuracy is required to place a measured centroid near the correct value. Isotope envelopes add elemental clues through M plus one and M plus two abundances. Chlorine and bromine patterns are especially diagnostic, while carbon-13 abundance scales roughly with carbon count for small molecules.
Require agreement among exact mass, isotope pattern, charge and source chemistry before accepting a formula.
What this chapter covers
- 01
Mass-to-charge ratio and ion notation
- 02
Nominal mass, exact mass and atomic average
- 03
Mass sufficiency and mass deficiency
- 04
Carbon-13 M plus one estimates
- 05
Chlorine and bromine M plus two patterns
- 06
Electron and chemical ionisation
- 07
Electrospray and MALDI source choice
Worked example · free
Recover neutral mass from a charged electrospray ion
- 3Use isotope spacing to identify the charge state.
- 5Multiply by charge and remove the two proton masses.
Key terms
- Mass-to-Charge Ratio
- The ion mass divided by its integer charge state, plotted as m/z.
- Exact Mass
- The sum of the exact masses of the specific isotopes in an ion.
- Isotope Envelope
- The cluster of peaks created by natural isotopic variants of one ionic composition.
- Electrospray Ionisation
- A soft solution-phase source that commonly creates multiply charged ions from polar molecules.
- MALDI
- Matrix-assisted laser desorption ionisation, a pulsed source often used for large molecules with TOF analysis.
Molecular Ions, Isotopes and Ionisation FAQ
How can isotope spacing reveal charge?
Adjacent isotope substitutions change ion mass by about one atomic mass unit. Dividing that change by charge gives the observed spacing, so one, one half and one third m/z correspond approximately to singly, doubly and triply charged ions.
Why is exact mass different from molecular weight?
Exact mass uses a specified isotopic composition, while a relative molecular mass based on elemental averages reflects natural abundance. Nominal mass uses integer mass numbers. Formula assignment requires the quantity that matches the measurement and convention.
How should I choose an ion source?
Start with sample phase, volatility, polarity and molecular size, then decide whether intact molecular mass or fragment information is the priority. The source must create ions compatible with the analyser and the analytical claim.
Exam move
Practise writing five common ion notations and converting each between neutral mass and m/z. Estimate M plus one abundance for several carbon counts, then check a chlorine or bromine envelope qualitatively. Build a source comparison that connects sample property to ion behaviour and analytical evidence rather than memorising names.
Working through Molecular Ions, Isotopes and Ionisation in CHEM6118? Sia is AskSia’s AI Chemistry tutor — ask any CHEM6118 Molecular Ions, Isotopes and Ionisation question and get a clear, step-by-step explanation grounded in how CHEM6118 is taught and assessed. Read this chapter free, then take your hardest questions to Sia.