PHYS5044 Chap.2 Calibration, Accuracy, Precision and Response Functions
Calibration, Accuracy, Precision and Response Functions
Calibration, Accuracy, Precision and Response Functions separates the two ways a reading can be wrong and shows how each one is reduced. Accuracy is the absolute uncertainty, the distance between a reading and the true value, and it is dominated by systematic offset and gain errors. Precision is the uncertainty in a change, set by random scatter between repeated readings.
The chapter explains why a task that follows a slow change needs precision while a task that checks an absolute limit needs accuracy. It then builds the two-point linear calibration, from the gain and offset of two references to the inverse rule that turns a voltage into a measurand, and shows how a third reference exposes curvature that only a polynomial or lookup table can remove.
Resolution and bit depth are kept apart from accuracy. The second half treats sensors as first-order systems: an exponential step response with time constant tau, a low-pass frequency response with corner frequency one over two pi tau, and the fraction of a fluctuation that survives. Antenna gain from aperture area and wavelength closes the chapter as a response that a datasheet quotes and a calculation can audit.
What this chapter covers
- 01
Accuracy as absolute uncertainty, precision as uncertainty in a change
- 02
Choosing which one an application needs
- 03
Two-point linear calibration and its inverse
- 04
Nonlinear response and multi-point calibration
- 05
First-order step response and time constant
- 06
Corner frequency and attenuation of a fluctuation
- 07
Antenna gain from aperture and wavelength
Worked example · free
Diagnose three instruments from their scatter on a target
- 1The centred but wide cluster is accurate on average but imprecise; averaging several readings, or narrowing the bandwidth, shrinks its scatter.
- 1The tight but shifted cluster is precise but biased; a calibration against a known reference removes the offset.
- 1The centred and tight cluster is both accurate and precise, so it needs only a periodic check against its reference.
- 2To follow a slow drift, the biased but tight gauge is already adequate, because a constant offset cancels when one reading is subtracted from the next, while scatter would hide small changes.
Key terms
- Systematic error
- A repeatable offset or gain error that shifts every reading the same way and survives averaging.
- Two-point calibration
- A calibration that fits a straight line through the readings at two known references, fixing gain and offset.
- Time constant
- Characteristic time of a first-order sensor: the time it takes to cover 63% of a sudden step in the measured quantity.
- Corner frequency
- The frequency at which a first-order response has fallen to 0.707 of its low-frequency value, one over two pi times the time constant.
- Aperture efficiency
- Measured antenna gain divided by the gain an ideal aperture of the same area would give.
Calibration, Accuracy, Precision and Response Functions FAQ
Is accuracy or precision more important?
It depends on the task. Following a change over time needs precision, because a constant offset cancels in the difference. Checking a value against an absolute limit needs accuracy, because any offset passes straight into the answer. State which one the application needs and give the reason in the same sentence.
Why test a calibration with a third point?
Two references always fit a straight line, so they cannot reveal curvature. A third reference in the middle of the range shows whether the response really is linear. If it disagrees by more than the scatter, a polynomial or lookup table is needed instead of a single gain and offset.
Does a higher bit depth make a reading more accurate?
No. More bits give smaller voltage steps, which improves resolution, but an offset or gain error is carried through every step unchanged. Accuracy improves through calibration against trusted references, while resolution improves through the converter, and the two should be quoted separately.
How do I get a time constant from a step response?
Find the fraction of the step reached at a known time, then solve the exponential: the remaining fraction equals e to the minus t over tau. At one time constant the output has covered 63% of the step, and after about 4.6 time constants it is within 1%.
What does the corner frequency tell me?
It marks where a first-order sensor passes 0.707 of a slow signal. Well above it the passed fraction falls in proportion to frequency, so a fluctuation at four times the corner keeps only about a quarter of its amplitude. Always state the fraction for the signal frequency in the question.
Exam move
Start with the two definitions and write one application for each where it matters most. Then practise two-point calibration by hand: compute a gain and offset from two references, invert the rule for a new reading, and test it against a third reference to judge whether curvature or scatter limits the result.
For response functions, take one time constant and produce the step response, the corner frequency, the settling time to 1% and the fraction of a faster fluctuation that survives. Sketch the exponential with labelled axes, since sketches carry marks. Finish with one antenna gain calculation in linear units and in dBi, and compare it with a datasheet figure to infer an aperture efficiency.
Close each session by naming, for every number you produced, whether it is an offset, a gain, a curvature, a scatter or a speed.
Working through Calibration, Accuracy, Precision and Response Functions in PHYS5044? Sia is AskSia’s AI Physics tutor — ask any PHYS5044 Calibration, Accuracy, Precision and Response Functions question and get a clear, step-by-step explanation grounded in how PHYS5044 is taught and assessed. Read this chapter free, then take your hardest questions to Sia.