BIOL10401 Chap.4 Dilutions, Calibration and Concentration
Dilutions, Calibration and Concentration
Dilutions, Calibration and Concentration use standards to turn instrument response into an unknown value. Known concentrations are measured under the same conditions as the unknown and plotted with concentration on the horizontal axis and absorbance on the vertical axis. The fitted relation gives the concentration inside the cuvette. A documented dilution factor is then restored once to recover the original sample.
Serial dilution factors multiply because each step retains a fraction of the previous tube. Unit conversion from mass concentration to molarity requires molecular mass after milligrams per millilitre has been recognised as grams per litre.
What this chapter covers
- 01
Standard series
- 02
Scatter plots
- 03
Best-fit relation
- 04
Unknown concentration
- 05
Serial dilution
- 06
Molar conversion
Worked example · free
Recover the original sample
- 2Solve 1.14 divided by 7.2 to obtain 0.1583 mg/mL in the cuvette.
- 1Multiply once by four to obtain 0.633 mg/mL in the original sample.
- 1Substitute the diluted value to check the measured absorbance.
Key terms
- Calibration curve
- A calibration curve relates known concentration to measured response.
- Serial dilution
- A serial dilution applies successive documented dilution steps.
- Molar concentration
- Molar concentration is amount of substance in moles per litre.
Dilutions, Calibration and Concentration FAQ
Why must standards span the unknown?
An unknown inside the validated standard range is interpolated. A value outside that range depends on extrapolation and may violate the response relation, so dilution and remeasurement are safer.
Do serial dilution factors add?
No. Each step acts on the preceding tube, so the fractions multiply. A two-fold step followed by a five-fold step gives a ten-fold total dilution.
When do I restore the dilution factor?
First use the curve to solve the concentration in the measured cuvette. Multiply that result by the total documented factor once to recover the original sample.
Exam move
Build calibration questions as a fixed sequence: identify standard axes, read or calculate the fitted relation, solve the concentration in the cuvette, restore any dilution once and convert units only after the sample stage is clear. Practise serial dilutions by recording sample volume, final volume and factor in a table. Audit graphs for axis labels, units, useful range and whether a forced origin is justified.
Recalculate one standard by hand to verify the spreadsheet formula. For molar conversion, write grams per litre above grams per mole so cancellation leaves moles per litre. Use substitution and direction checks to catch values that look plausible but belong to the wrong tube. Start every calibration problem by defining x and y in words and units.
Known standard concentration normally belongs on the horizontal axis and measured response on the vertical axis. For y = mx + b, substitute the unknown sample response for y and solve x = (y - b)/m. Do not silently discard the intercept or force the fit through zero unless the preparation and blanking justify that model.
Check whether the unknown response lies inside the standard range; interpolation is supported by measured standards, whereas extrapolation assumes the relation continues beyond them. Record fit quality without treating a strong coefficient as proof that preparation errors are absent. Separate the cuvette result from the original sample.
If a sample was diluted before measurement, the calibration equation first returns concentration in the measured dilution. Multiply by the reciprocal dilution only once to recover the previous stage. Draw a chain of labelled tubes and write concentration beneath each one. In a serial dilution, each transfer acts on the concentration of the tube immediately before it, so overall dilution is the product of step factors.
A table containing sample volume, diluent volume, final volume and step factor makes an omitted or doubled correction visible. Mix every tube before transferring onward because exact arithmetic cannot repair a non-representative aliquot. Build dimensional checks into the solution rather than adding them afterward.
Convert microlitres to millilitres before using a plated or measured volume, and distinguish micrograms per millilitre from milligrams per litre even when their numerical values coincide. To convert mass concentration to molarity, divide grams per litre by grams per mole and verify that moles per litre remain.
For a predicted response, substitute a known standard concentration back into the fitted equation and compare with the observed point. For an unknown, ask whether concentration direction matches response direction when slope is positive. These checks expose a plausible number that belongs to the wrong axis, wrong tube or wrong unit.
Build one complete practice chain from stock preparation to serial dilution, instrument response, fitted equation and recovered original concentration. Mark each point where a factor is introduced so the final answer contains no hidden or duplicated correction.
Working through Dilutions, Calibration and Concentration in BIOL10401? Sia is AskSia’s AI Biology tutor — ask any BIOL10401 Dilutions, Calibration and Concentration question and get a clear, step-by-step explanation grounded in how BIOL10401 is taught and assessed. Read this chapter free, then take your hardest questions to Sia.