MRI Signal: Spin, Precession and Relaxation

This chapter explains where the MRI signal comes from. Hydrogen nuclei in water behave as tiny magnets that align with the main field and precess at the Larmor frequency. A radio-frequency pulse at that frequency tips the net magnetisation, and two relaxation processes, T1 recovery along the field and T2 decay across it, set how bright each tissue appears.

A worked example ranks four tissues using the lecture's own T1 and T2 table. The physics here underpins Chapters 3 and 4, because pulse-sequence timing and diffusion weighting are both ways of choosing which part of this relaxation behaviour the image shows. Expect both short calculations, such as a transmit frequency or a recovery fraction, and short explanations of why a tissue looks bright or dark.

In this chapter

What this chapter covers

  • 01

    Nuclei with spin and charge, and why hydrogen dominates

  • 02

    Net magnetisation and alignment with B0

  • 03

    Precession and the Larmor equation

  • 04

    RF excitation and the emitted signal

  • 05

    T1 recovery and the 63% rule

  • 06

    T2 decay and loss of phase coherence

  • 07

    Tissue T1 and T2 values and brightness

Worked example · free

Larmor frequency and longitudinal recovery

Q [3 marks]. A 3.0 T scanner images white matter with T1 = 680 ms. Find the hydrogen excitation frequency, then the fraction of longitudinal magnetisation recovered 1,360 ms after a 90° pulse. Marks shown follow our own practice weighting, not an official allocation.
Net magnetisation before and after excitation.Net magnetisation before and after excitation. At rest the net vector M points along B0; a 90° RF pulse tips it into the transverse plane, while each proton precesses around B0 at the Larmor frequency shown on the right.B0 (z axis)M at restM after 90° pulsetransverse planeprecession at ωone proton
Net magnetisation before and after excitation. At rest the net vector M points along B0; a 90° RF pulse tips it into the transverse plane, while each proton precesses around B0 at the Larmor frequency shown on the right.
  • 1ω = γB0 = 42.58 × 3.0 = 127.74 MHz.
  • 11,360 ms is 2 × T1, so Mz/M0 = 1 − e^(−2).
  • 11 − 0.135 = 0.865, so about 86.5% has recovered.
127.74 MHz, and about 86.5% recovery after two T1 periods. The frequency scales linearly with the field, so the same tissue at 1.5 T would be excited at half this value, while its recovery fraction depends only on time measured in units of T1. That separation is the key idea of the chapter.
Sia tip — Express the waiting time as a multiple of T1 first; one T1 gives 63%, two give 86.5%, five give almost full recovery.
Glossary

Key terms

Larmor frequency
The Larmor frequency is the rate at which a nucleus precesses around B0, equal to γ times the field strength.
T1 relaxation
T1 relaxation is the exponential regrowth of longitudinal magnetisation along B0 after excitation.
T2 relaxation
T2 relaxation is the decay of transverse magnetisation as precessing protons lose phase coherence.
Gyromagnetic ratio
The gyromagnetic ratio is a nucleus's magnetic moment divided by its angular momentum, 42.58 MHz per tesla for hydrogen.
Net magnetisation
Net magnetisation is the vector sum of all proton magnetic moments in a voxel, aligned with B0 at rest.
Precession
Precession is the rotation of a proton's magnetic moment around the direction of the main field B0.
FAQ

MRI Signal: Spin, Precession and Relaxation FAQ

Why does MRI image hydrogen rather than other nuclei?

Hydrogen is the only major element in tissue that is MR sensitive, and it is also the most abundant atom in the body, mostly in water. Other MR-capable nuclei such as carbon-13 or sodium-23 are far scarcer, so their signal is weak.

What happens if the RF pulse is not at the Larmor frequency?

The protons absorb almost no energy because resonance only occurs at the precession frequency. The net magnetisation is barely tipped, so little transverse signal is produced and the receive coils pick up almost nothing.

Why is water dark on T1-weighted images but bright on T2-weighted images?

Water has long T1 and long T2, both 3000 ms in the lecture table. Its slow recovery makes it dark when T1 differences dominate, while its slow decay keeps signal high when the echo is taken late and T2 dominates.

What is the difference between in phase and out of phase?

Protons precessing together are in phase and their transverse components add up to a measurable signal. When they precess separately they are out of phase, their components cancel, and the signal fades; that loss of coherence is what T2 relaxation describes.

Why does MRI not use ionising radiation?

The signal comes from radio-frequency energy absorbed and re-emitted by hydrogen nuclei in a magnetic field, not from X-rays or gamma rays. That is why the lecture lists MRI as safe for people vulnerable to radiation, such as pregnant women and babies.

How strong is a 3 tesla MRI field?

A tesla is one newton per ampere-metre, so in a 3.0 T field a wire carrying 1 A at right angles feels 3.0 N per metre. The lecture compares this with about 5 mT for a refrigerator magnet and 30 to 70 microtesla for the Earth's field.

Study strategy

Exam move

Learn the Larmor equation with the value 42.58 MHz per tesla and practise it at 1.5, 3 and 7 T. Then memorise the T1 and T2 table and the two orderings, noting where the slide's curve exercise and the table disagree about fat. Sketch recovery and decay curves for two tissues and mark where they cross 63% and 37%.

Finally, explain in two sentences why a strong field increases the transmit frequency but not the T1 ordering of tissues, and practise converting between a recovery fraction and a waiting time. When you revise, keep the two axes separate in every sentence: T1 always refers to the z component regrowing, T2 always to the x-y component fading.

Working through MRI Signal: Spin, Precession and Relaxation in ELEC5622? Sia is AskSia’s AI Electrical Engineering tutor — ask any ELEC5622 MRI Signal: Spin, Precession and Relaxation question and get a clear, step-by-step explanation grounded in how ELEC5622 is taught and assessed. Read this chapter free, then take your hardest questions to Sia.

A+Everything unlocked
Unlocks this Bible + all 196 of your University of Sydney subjects - and 1,000+ Bibles across every Australian university.
Sia - your ELEC5622 tutor, unlimited, worked the way the exam marks it
The full 4-page Bible + practice bank with worked solutions
Chrome extension - sync your LMS so Sia knows your deadlines
Bilingual EN / Chinese on every Bible and every Sia answer
$0.99 Trial
30-day money-back · cancel in one tap · how it works